Categories

Viewed products

LHT-Series Impact Devices

LHT-Series Impact Devices

LHT-Series Impact Devices

Separate impact devices are available to suit the range of portable LEEB hardness testers. They allow a single testing unit to be utilized for an expanded range of non standard applications including small, lightweight parts, recessed grooves, notches and large, heavy rough castings by changing the separate impact device to one that is suitable for the requirements.

Availability: In stock, Ships within 1 business day

$ 495.00 + GST

Add to cart

More info

The standard supply with a portable hardness tester, to cover most uses, is with a D-Type impact device. For a wider range of testing abilities and for certain material types, weights and thicknesses, it may be neccesary to change to a different type of impact device suitable to the sample being measured. 

These impact devices are interchangable. For use with an additional impact device the LHT-200 has auto recognition which detects that a different device is connected and applies relevant settings which become visible when the different device is connected.

Different devices and their applications are as follows:

 

D Type -  Used in 90% of applications the d-type impact device is the standard device supplied.  It works on material that is greater than 5mm of thickness and good for general use.

C Type - Suitable for light and small work pieces from about 1mm of thickness depending on the material.  This impact device is also better for testing materials with surface hardened layers as the C-type applies a lower force when testing.

DC Type - Designed for reduced access applications.  Shorter than the standard unit it has all the functionality of the D-type impact device. 

DL Type - Designed to gain access in narrow areas that other gauges cannot.  This unit can be used for testing gear teeth, long channels, holes and recesses.

G Type - Designed for rough, heavy objects.  It has a heavy duty device that works on an uneven, unpolished surface such as castings.

If you are unsure on the most suitable device for your application, please contact us to discuss.

 

Impact Device D/DC/DL G C
Impacting Energy 11mJ 90mJ 2.7mJ
Mass of Impact body 5.5g/5.5g/7.2g 20g 3.0g
Diameter 20/20/6mm 30mm 20mm
Length 147/86/202mm 255mm 141mm
Weight 75/50/60g 250g 75g
Max. Hardness of workpiece 940/940/950HV 650HV 1000HV
Surface Roughness of workpiece ≤1.6µm ≤6.3µm ≤0.4µm
Min weight of workpiece (direct) >5Kg >15Kg >1.5Kg
Min weight of workpiece (support) 2~5Kg 5~15Kg 0.5~1.5Kg
Min weight of workpiece (coupled) 0.05Kg~2 0.5~5Kg 0.02~0.5Kg
Min thickness of workpiece (coupled) 5mm 10mm 1mm
Min thickness of hardened layer 0.8mm 1.2mm 0.2mm
Hardness of spherical test tip 1600HV
Material of spherical test tip Tungsten Carbide


DIMENSIONS: